| Specification Parameter | Technical Details |
|---|---|
| Model / Series | CC-W02 Multiparameter Water Quality Analyzer (CC-W Series) |
| System Architecture | Multi-channel Intelligent Analyzer Box with a maximum 5-parameter synchronization loop system |
| Monitored Key Variables Matrix | Temperature, pH, ORP, Conductivity (EC), TDS, Salinity, Dissolved Oxygen (DO), Residual Chlorine, Dissolved Ozone, Turbidity, TSS, Sludge Concentration, COD, Chlorophyll, and BGA |
| Compatible Digital Sensors Matrix Specs | |
| Temp. Sensor | 0°C – 60°C | Principle: Thermal method | Res: 0.01°C | Accuracy: ±0.5°C |
| pH Sensor | 0 – 14 pH | Principle: Glass electrode method | Res: 0.01 pH | Accuracy: ±0.1 pH |
| ORP Sensor | -2000mV – +2000mV (Supports -1999~+1999 mV scale) | Principle: Glass electrode method | Res: 1mV | Accuracy: ±2mV |
| Conductivity Sensor (EC) | 0 – 20000 µS/cm (K=1) (Supports 0-20 mS/cm customizable scale) | Principle: Conductivity cell method | Res: 0.1 µS/cm or 0.01 mS/cm | Accuracy: ±1.5% FS |
| DO Sensor (Dissolved Oxygen) | 0 – 20 mg/L (0-20 ppm), 0 – 200% | Principle: Fluorescence method | Res: 0.1% / 0.01 mg/L | Accuracy: ±3% or ±0.3 mg/L |
| Turbidity Sensor | 0.1 – 1000 NTU (Supports 0-500 NTU limit scale) | Principle: Light scattering | Res: 0.01 – 0.1 NTU | Accuracy: ±5% or 0.3 NTU |
| SS Sensor (Suspended Solids / TSS) | 0.01-20000 mg/L | 0.01-45000 mg/L | 0.01-120000 mg/L | Principle: Light scattering | Res: 0.01 – 1 mg/L | Accuracy: ≤ ±5% measured (Depends on sludge homogeneity) |
| Chlorophyll Sensor (Chl-a) | 0 – 500 µg/L | Principle: Fluorescence method | Res: 0.01 µg/L | Accuracy: ±5% of signal level of 1 ppb rhodamine B dye | Linearity: R² > 0.999 |
| BGA Sensor (Blue-green Algae) | 0 – 300,000 cells/mL | Principle: Fluorescence method | Res: 20 cells/mL | Accuracy: ±10% of signal level of 1 ppb rhodamine B dye | Linearity: R² > 0.999 |
| COD Sensor | 0 – 500 mg/L COD (5mm) | Principle: UV absorption method | Res: 0.01 mg/L | Accuracy: ±5% FS |
| System & Hardware Configurations | |
| Signal Output Configuration | RS485 (Modbus-RTU) communication interface |
| Power Supply Voltage | 12 VDC |
| Cable Length | Fully customizable configurations available upon request (Optional) |
| Tailored OEM/ODM Capability | • Brand Customization: Custom silk-screen or laser-engraved logo branding available. • Packaging Options: Private-label custom colored packaging designs. • Technical Adaptation: Fully support customizable measurement factors, sensing ranges, hardware resolutions, and multi-channel integration configurations. |
| Applications | Multi-parameter online water tracking across industrial wastewater treatment plants, municipal water grids, environmental river telemetry, aquaculture facility control, secondary water supply systems, and chemical process control. |
To request the complete PDF technical specifications, please submit your inquiry below.