CC-E-G12 Multi-Parameter Air Quality Sensor

• CC-E-G12 compact outdoor multi-parameter air quality sensor for PM2.5, PM10, TSP, SO2, NO2, O3, and CO monitoring
Specification Parameter Technical Details
Model / Series CC-E-G12 Multi‑Parameter Air Quality Sensor
Key Measurements Matrix Modular configurable: optional gas channels + particulate matter (PM2.5, PM10, TSP)
Air Quality: Particulate Matter
Particle Size Channels0.3~1.0 μm; 1.0~2.5 μm; 2.5~10 μm
Particle Counting Efficiency50% @ 0.3 μm; 98% @ > 0.5 μm
PM2.5 Effective Range0~1000 µg/m³ | Max Range: ≥ 5000 µg/m³
PM2.5 Measurement Consistency±10 µg/m³ @ 0~100 µg/m³; ±10% @ 100~1000 µg/m³
PM10 Effective Range0~1000 µg/m³ | Max Range: ≥ 5000 µg/m³
PM10 Measurement Consistency±15 µg/m³ @ 0~100 µg/m³; ±15% @ 100~1000 µg/m³
TSP Effective Range0~2000 µg/m³ | Max Range: ≥ 10000 µg/m³
Mass Concentration Resolution (Particles)1 µg/m³
Air Quality: Optional Gas Sensors Library (All gases are modular optional)
Sulfur Dioxide (SO₂) – 0~5 ppmRange: 0~5 ppm | Resolution: 0.01 ppm | Repeatability: < 2% | Accuracy: ±5% F.S | T90: ≤ 30 s
Sulfur Dioxide (SO₂) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Repeatability: < 2% | Accuracy: ±5% F.S | T90: ≤ 30 s
Sulfur Dioxide (SO₂) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Sulfur Dioxide (SO₂) – 0~1000 ppmRange: 0~1000 ppm | Resolution: 1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Nitrogen Dioxide (NO₂) – 0~5 ppmRange: 0~5 ppm | Resolution: 0.01 ppm | Repeatability: < 2% | Accuracy: ±5% F.S | T90: ≤ 30 s
Nitrogen Dioxide (NO₂) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Repeatability: < 2% | Accuracy: ±5% F.S | T90: ≤ 30 s
Nitrogen Dioxide (NO₂) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Ozone (O₃) – 0~5 ppmRange: 0~5 ppm | Resolution: 0.01 ppm | Repeatability: < 2% | Accuracy: ±5% F.S | T90: ≤ 30 s
Ozone (O₃) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Repeatability: < 2% | Accuracy: ±5% F.S | T90: ≤ 30 s
Carbon Monoxide (CO) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Repeatability: < 2% | Accuracy: ±5% F.S | T90: ≤ 25 s
Carbon Monoxide (CO) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 25 s
Carbon Monoxide (CO) – 0~500 ppmRange: 0~500 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 25 s
Carbon Monoxide (CO) – 0~1000 ppmRange: 0~1000 ppm | Resolution: 1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 25 s
Combustible Gas – 0~100%LELRange: 0~100%LEL | Resolution: 0.01%LEL | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Methane (CH₄) – 0~100%LELRange: 0~100%LEL | Resolution: 0.01%LEL | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Methane (CH₄) – 0~100%VolRange: 0~100%Vol | Resolution: 0.01%Vol | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Acetylene (C₂H₂) – 0~100%LELRange: 0~100%LEL | Resolution: 0.01%LEL | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Ethylene (C₂H₄) – 0~100%LELRange: 0~100%LEL | Resolution: 0.01%LEL | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Ethane (C₂H₆) – 0~100%LELRange: 0~100%LEL | Resolution: 0.01%LEL | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Hydrogen (H₂) – 0~1000 ppmRange: 0~1000 ppm | Resolution: 1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Hydrogen (H₂) – 0~100%LELRange: 0~100%LEL | Resolution: 0.01%LEL | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Oxygen (O₂) – 0~30%VolRange: 0~30%Vol | Resolution: 0.1%Vol | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Oxygen (O₂) – 0~100%VolRange: 0~100%Vol | Resolution: 1%Vol | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Hydrogen Sulfide (H₂S) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Hydrogen Sulfide (H₂S) – 0~50 ppmRange: 0~50 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Hydrogen Sulfide (H₂S) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Chlorine (Cl₂) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Chlorine (Cl₂) – 0~20 ppmRange: 0~20 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Ammonia (NH₃) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Ammonia (NH₃) – 0~50 ppmRange: 0~50 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Ammonia (NH₃) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Hydrogen Fluoride (HF) – 0~5 ppmRange: 0~5 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Hydrogen Fluoride (HF) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Hydrogen Chloride (HCl) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Hydrogen Chloride (HCl) – 0~20 ppmRange: 0~20 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Phosphine (PH₃) – 0~5 ppmRange: 0~5 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Phosphine (PH₃) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Arsine (AsH₃) – 0~0.5 ppmRange: 0~0.5 ppm | Resolution: 0.001 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Arsine (AsH₃) – 0~1 ppmRange: 0~1 ppm | Resolution: 0.001 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Silane (SiH₄) – 0~5 ppmRange: 0~5 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Silane (SiH₄) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Nitric Oxide (NO) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Nitric Oxide (NO) – 0~500 ppmRange: 0~500 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Nitrous Oxide (N₂O) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Formaldehyde (HCHO) – 0~5 ppmRange: 0~5 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Formaldehyde (HCHO) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Methanol (CH₃OH) – 0~50 ppmRange: 0~50 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Methanol (CH₃OH) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Ethanol (C₂H₅OH) – 0~50 ppmRange: 0~50 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Ethanol (C₂H₅OH) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Acetone – 0~50 ppmRange: 0~50 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Acetone – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
VOC (PID) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
VOC (PID) – 0~50 ppmRange: 0~50 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
VOC (PID) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
VOC (PID) – 0~1000 ppmRange: 0~1000 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
VOC (PID) – 0~60000 ppmRange: 0~60000 ppm | Resolution: 1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Propane (C₃H₈) – 0~100%LELRange: 0~100%LEL | Resolution: 0.01%LEL | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Butane (C₄H₁₀) – 0~100%LELRange: 0~100%LEL | Resolution: 0.01%LEL | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Isobutane – 0~100%LELRange: 0~100%LEL | Resolution: 0.01%LEL | Accuracy: ≤ ±15% F.S | T90: ≤ 10 s
Toluene (C₇H₈) – 0~50 ppmRange: 0~50 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Toluene (C₇H₈) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Xylene (C₈H₁₀) – 0~50 ppmRange: 0~50 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Xylene (C₈H₁₀) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Ethylbenzene – 0~50 ppmRange: 0~50 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Ethylbenzene – 0~100 ppmRange: 0~100 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Hydrogen Cyanide (HCN) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Hydrogen Cyanide (HCN) – 0~20 ppmRange: 0~20 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Bromine (Br₂) – 0~10 ppmRange: 0~10 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Bromine (Br₂) – 0~20 ppmRange: 0~20 ppm | Resolution: 0.01 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Oxygen Difluoride (OF₂) – 0~1 ppmRange: 0~1 ppm | Resolution: 0.001 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Sulfur Hexafluoride (SF₆) – 0~100 ppmRange: 0~100 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Sulfur Hexafluoride (SF₆) – 0~1000 ppmRange: 0~1000 ppm | Resolution: 0.1 ppm | Accuracy: ≤ ±15% F.S | T90: ≤ 30 s
Other Custom GasesMany more detectable gases available; contact our engineering team for gas types, measurement ranges and specifications not listed above.
System & Hardware Configurations
Output Signal Configuration RS485 (Modbus-RTU) serial interface
Power Supply Voltage 12 ~ 24 VDC
Cable Length Fully customizable configurations available upon request (Optional)
Tailored OEM/ODM Capability • Brand Customization: Custom silk-screen or laser-engraved logo branding available.
• Packaging Options: Private-label custom colored packaging designs.
• Technical Adaptation: Fully customizable gas combinations, measurement ranges, sensing channels, resolution rates, and output signal configurations.
Applications Urban ambient air quality monitoring, industrial park emission supervision, factory boundary monitoring, mining gas detection, smart city environmental stations, road-side air telemetry and environmental research networks.
To request the complete PDF technical specifications, please submit your inquiry below.

Product Overview

The CHAOSENSOR CC‑E‑G12 Multi‑Parameter Air Quality Sensor is a fully modular outdoor‑grade environmental sensing unit combining electrochemical, laser‑scattering, NDIR and PID detection principles. Designed for flexible channel customization, this multi‑parameter air quality sensor supports optional gas modules plus particulate‑matter (PM2.5, PM10, TSP) measurement for smart‑city environmental IoT nodes, industrial‑park boundary monitoring, roadside air telemetry and distributed ambient‑quality observation networks. If you are evaluating alternative air‑quality hardware solutions, please review our CHAOSENSOR CC‑E‑AQM Air Quality Monitoring Station (active pump‑suction cabinet‑type station) and our compact CHAOSENSOR CC‑E‑G08 Multi‑Parameter Air Quality Sensor for cost‑optimized deployment scenarios.

Core Working Principle

The CC‑E‑G12 integrates multiple mature sensing technologies for comprehensive air‑quality measurement:

  • Electro‑Chemical Gas Sensors: Detect toxic, combustible and industrial gases via electrochemical cell reaction, delivering fast T90 response for target gaseous pollutants.
  • NDIR (Non‑Dispersive Infrared): Used for high‑stability measurement of infrared‑absorbing gases such as CO₂ and CH₄.
  • Photo‑Ionization Detector (PID): For broad‑range VOC / TVOC organic‑vapor concentration monitoring.
  • Laser‑Scattering Particle Detector: Counts and converts light‑scattering signals into mass concentration for PM2.5, PM10 and TSP particulate‑matter fractions.

All sensor‑module outputs are processed by the on‑board microcontroller and transmitted via industry‑standard RS485 Modbus‑RTU for data‑logger, PLC, SCADA or IoT‑gateway integration. Gas channels are fully modular; users select only required gas species and measurement ranges for each project.

Key Features & Engineering Advantages

  • Fully Modular Sensing Matrix: All gas channels are optional plug‑in modules. Configure exact gas species, measurement ranges and resolution to match site requirements without paying for unused sensors.
  • Broad Gas‑Species Coverage: Supports combustible gases, common toxic‑industrial gases, VOC‑family organics and specialty rare‑process gases, covering most environmental‑monitoring and factory‑perimeter‑detection use‑cases.
  • Integrated Particulate‑Matter Measurement: Built‑in PM2.5 / PM10 / TSP laser‑scattering particle‑monitoring function, eliminating extra particle‑sensor hardware for combined gas‑and‑dust monitoring.
  • Industrial‑Grade Communication: Standard RS485 Modbus‑RTU digital interface, well‑suited for multi‑drop bus networking over long‑field‑cable runs with industrial data‑acquisition systems.
  • Wide‑Range Power Input: Operates on 12‑24 VDC, compatible with solar‑battery off‑grid deployments as well as mains‑powered weather‑station installations.
  • Complete OEM / ODM Customization: Supports custom laser‑engraved branding, private‑label packaging, adjusted output‑scaling and tailored channel‑combinations for large‑volume tender‑project requirements.

Sensor Selection Guide: CC‑E‑G12 vs CC‑E‑AQM vs CC‑E‑G08

CHAOSENSOR offers three major multi‑parameter air‑quality product families for different project‑deployment strategies:

  • CHAOSENSOR CC‑E‑G12 Multi‑Parameter Air Quality Sensor: Modular outdoor louver‑shield‑type sensor unit. Best for distributed dense‑node monitoring where you need flexible gas‑channel mixing plus built‑in PM measurement; passive ambient‑air intake without active pump‑suction.
  • CHAOSENSOR CC‑E‑AQM Air Quality Monitoring Station: Cabinet‑type station equipped with active pump‑suction sampling‑loop. Recommended for high‑grade industrial‑park perimeter monitoring that requires controlled, consistent air‑sample aspiration conditions.
  • CHAOSENSOR CC‑E‑G08 Multi‑Parameter Air Quality Sensor: Compact‑form multi‑gas sensor without integrated particulate‑matter channels, for cost‑sensitive projects requiring only gaseous‑pollutant measurement.

Choose CC‑E‑G12 when your project requires both configurable‑modular gas monitoring and integrated PM2.5 / PM10 / TSP particulate‑matter measurement for distributed outdoor environmental‑monitoring networks.

Typical Industrial Applications

  • Smart‑City Distributed Environmental IoT Networks: Large‑scale grid‑layout air‑quality sensing for urban environmental assessment, community‑level pollution mapping and municipal‑traffic‑emission research.
  • Industrial‑Park & Factory‑Boundary Air‑Quality Supervision: Perimeter monitoring for chemical‑industrial zones, detecting fugitive toxic‑gas emissions together with particulate‑pollutant levels.
  • Road‑Side & Traffic‑Related Air‑Telemetry: Deployed along highways and arterial roads to capture traffic‑related gaseous‑pollutant and PM‑concentration data.
  • Environmental‑Impact Assessment & Research Projects: Field‑deployable multi‑parameter sensor for short‑term and long‑term ambient‑air‑quality investigation tasks.
  • Comprehensive Automatic Weather‑Station Retrofit: Add multi‑gas and particulate‑matter channels to existing AWS installations for expanded atmospheric‑observation capability.

Installation & Field‑Deployment Guidelines

Follow these installation best‑practices to preserve CC‑E‑G12 measurement‑performance in outdoor field‑conditions:

  • Mounting Position: Install in open representative ambient‑air location. Avoid direct exhaust outlets, building‑wall jet‑flows, obstructions and heavy dust‑deposition hot‑spots that distort sampled‑air composition.
  • Mechanical Mount: Secure firmly to rigid meteorological‑pole or bracket. Prevent excessive vibration induced by strong‑wind gusts.
  • Wiring Specification: Use shielded cable for RS485 communication. Strictly follow wiring diagram for 12‑24 VDC power‑supply polarity to avoid hardware damage.
  • Routine Maintenance: Periodically inspect louver radiation‑shield surface; clear accumulated dust, insect‑debris and bird‑droppings to maintain unobstructed ambient‑air exchange.
  • Surge‑Protection Recommendation: For lightning‑prone outdoor sites, install external surge‑protection devices on power‑and‑communication lines to protect sensor and backend data‑acquisition hardware.

Frequently Asked Questions (FAQ)

1. What is the key difference between CC‑E‑G12 and CC‑E‑AQM Air Quality Monitoring Station?
CC‑E‑G12 is a passive‑intake louver‑shield‑type modular sensor with built‑in PM channels. The CC‑E‑AQM station adopts an active pump‑suction sampling‑cabinet design, providing controlled aspiration sampling for high‑demanding industrial‑perimeter monitoring scenarios.

2. Should I select CC‑E‑G12 or CC‑E‑G08 Multi‑Parameter Air Quality Sensor?
Pick CC‑E‑G12 if you need combined gas plus particulate‑matter (PM2.5 / PM10 / TSP) measurement. Select CC‑E‑G08 if you only require gas‑pollutant monitoring and can omit particulate‑matter channels for better project‑cost control.

3. Are all gas‑detection channels on CC‑E‑G12 optional?
Yes. Every gas‑species module is modular‑optional. You can select exact gas types and corresponding measurement‑ranges according to your monitoring‑target; you are not forced to install the full gas‑matrix. PM particulate‑matter measurement is integrated as the base measurement function of CC‑E‑G12.

4. What communication protocols does CC‑E‑G12 support?
Standard RS485 hardware implements Modbus‑RTU protocol, compatible with mainstream industrial data‑loggers, PLCs, RTUs and environmental‑monitoring IoT‑gateway equipment.

5. Can CHAOSENSOR provide OEM / ODM custom‑configuration for CC‑E‑G12?
Yes. Custom laser‑engraved logo branding, private‑label packaging, customized gas‑channel combinations, adjusted measurement‑ranges and special output‑scaling parameters are supported for bulk‑order and project‑tender requirements.

This field is required.
This field is required.
This field is required.
This field is required.
This field is required.